BRIEF COMMUNICATION
Year : 2013  |  Volume : 17  |  Issue : 7  |  Page : 289-291

Erroneous glucose recordings while using mutant variant of quinoprotein glucose dehydrogenase glucometer in a child with galactosemia


1 Department of Endocrinology, St. John's Medical College, Bangalore, Karnataka, India
2 Department of Pediatrics, St. John's Medical College, Bangalore, Karnataka, India

Date of Web Publication11-Oct-2013

Correspondence Address:
Vivek Mathew
Department of Endocrinology, St. Johns Medical College Hospital, Sarjapur Road, Bangalore - 560 076, Karnataka
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/2230-8210.119616

Rights and Permissions
   Abstract 

We report a 2-month-old child with galactosemia and falsely high glucose readings with a glucometer using mutant variant of quinoprotein glucose dehydrogenase (MutQ-GDH) chemistry. Potentially fatal hypoglycemia could have been induced in the child if insulin infusion had been initiated as per glycemic management protocol. Even though, the product information with the glucometer carries warning regarding interference by high galactose levels, the awareness regarding this interaction is generally poor in many practice settings. Although, false readings have been reported with glucose dehydrogenase pyrroloquinoline quinone (GDH-PQQ) glucometers, to our knowledge this is the first case report of a falsely high glucose reading due to high galactose in a proven case of galactosemia with a glucometer using the MutQ-GDH chemistry (a modified GDH-PQQ chemistry). Our experience has prompted us to write this case report and we suggest avoiding these glucometers in neonates and infants when a metabolic disease is suspected.

Keywords: Galactosemia, glucose dehydrogenase pyrroloquinoline quinone glucometers, hypoglycemia, mutant variant of quinoprotein glucose dehydrogenase chemistry, neonate


How to cite this article:
Mathew V, Ramakrishnan A, Srinivasan R, Sushma K, Bantwal G, Ayyar V. Erroneous glucose recordings while using mutant variant of quinoprotein glucose dehydrogenase glucometer in a child with galactosemia. Indian J Endocr Metab 2013;17, Suppl S1:289-91

How to cite this URL:
Mathew V, Ramakrishnan A, Srinivasan R, Sushma K, Bantwal G, Ayyar V. Erroneous glucose recordings while using mutant variant of quinoprotein glucose dehydrogenase glucometer in a child with galactosemia. Indian J Endocr Metab [serial online] 2013 [cited 2019 May 25];17, Suppl S1:289-91. Available from: http://www.ijem.in/text.asp?2013/17/7/289/119616


   Introduction Top


Glucometers, which employ the glucose dehydrogenase pyrroloquinoline quinone (GDH-PQQ) are known to produce falsely high recordings of capillary glucose in the presence of interfering substances like maltose and galactose. [1] The food and drug administration (FDA) did recommend that these meters should not be used in hospital settings. [2] Following this public health notification GDH-PQQ glucometers have been supplanted by glucometers using mutant variant of quinoprotein glucose dehydrogenase (MutQ-GDH) chemistry which is a maltose independent advanced chemistry. We report a child who had falsely high capillary glucose values with a MutQ-GDH glucometer.

The child was later proven to have galactosemia and the falsely high capillary glucose was due to a cross reaction with galactose.


   Case Report Top


A 2-month-old male child was admitted to our hospital for jaundice and failure to thrive. The child was the first child born of a third degree consanguineous marriage. His birth weight was 3.5 kg. At 1 month of age, the child was noticed to have a yellowish green discoloration of the conjunctiva. The parents reported feeding difficulties and poor weight gain. His current weight was 2.8 kg. He was anemic, icteric and had hepatomegaly. The liver function tests revealed a cholestatic pattern. Admission arterial blood gas was suggestive of compensated metabolic acidosis with a high anion gap. His urine examination revealed ketonuria. Upon admission, random capillary glucose was performed, which revealed a value of 310 mg/dL [Table 1].
Table 1: Lab results of the patient on the first day of hospitalization

Click here to view


Subsequent readings taken every hour for the next 6 h were above 250 mg/dL. Capillary glucose testing was performed using Accu-Check® Performa (Roche diagnostics, Mannheim, Germany) glucometer. In view of suspected diabetic ketoacidosis, an endocrinology consultation was sought and hydration and insulin infusion were considered. However, the corresponding random venous plasma glucose was 67 mg/dL (performed using glucose oxidase method) and there was an absence of glucosuria. Hence, it was decided to withhold insulin infusion and potentially fatal hypoglycemia was avoided. In view of a possible cross reaction of high galactose with the glucometer, which employs a MutQ-GDH chemistry further testing was not performed using the same glucometer. A Benedict's test of urine showed the presence of reducing sugars. A quantitative test for galactose-1-phosphate uridyl transferase in whole blood showed a value of 3.85 U/g Hb (normal range 11-41.0 U/g Hb).

Quantitative test for galactose 1-phosphate in red blood cell was 1.8 mg/dL (normal range 0-1 mg/dL). A liver biopsy revealed macrovesicular steatosis with bile duct proliferation and a fibrosis score of 4/6 [Figure 1].
Figure 1: Liver biopsy showing macrovesicular steatosis (white arrow) and bileduct proliferation (black arrow)

Click here to view


The child was started on a galactose free formula. He showed steady improvement with resolution of jaundice and weighed 4.0 kg on repeat examination 1 month later.


   Discussion Top


Galactosemia is a rare condition with a prevalence of 1 in 60,000 live births among caucasians. Diagnosis of galactosemia is usually made by a new born screening test in western countries. In India, new born screening for galactosemia is not widely prevalent.

A prospective screening study for galactosemia in 18,000 newborns in India identified a single case of galactosemia with an estimated prevalence of 1:10,300. [3] However, the exact prevalence of galactosemia in India is still not clear.

Point of care testing devices like glucometers are increasingly being used in hospital care settings. Hence, it is important that the practicing clinician is aware of the technical deficiencies associated with these devices. There are mainly four sources of errors associated with glucometers. Pre-analytical errors include damage to the test strip or incomplete deposition of reagent on the test strip. Analytical errors deal with the technique of testing and includes insufficient or excessive blood loading, miscoding, calibration errors and those arising from contamination. Post-analytical errors result from misleading of the result display. Operator dependent errors can be divided into two, namely those involving glucometers using glucose oxidase chemistry and those involving GDH-PQQ chemistry.

Hypertriglyceridemia, hyperuricemia, hypoxia and acetaminophen can interfere with glucometers using the glucose oxidase chemistry. Diet or medication containing excessive galactose, maltose, xylose and vitamin C can interfere with glucose dehydrogenase (GDH) biosensor. [4] A previous analysis of the food and drug administration's (FDA) manufacturer and user facility device experience (MAUDE) data base for GDH-PQQ glucometer has listed two cases of severe hypoglycemia (non-fatal) in children with galactosemia and these events happened despite warnings by both manufacturers and regulatory authorities. [1] To our knowledge, this is the first clinical case report of a falsely high glucose reading due to high galactose in a proven case of galactosemia with a glucometer using the MutQ-GDH chemistry (a modified GDH-PQQ chemistry). The package insert of the glucometer cautions that galactose levels greater than 15 mg/dL (0.83 mmol/L) can cause overestimation of glucose values. According to a previous biochemical study galactose levels more than 39 mg/dL (2.2 mmol/L) had significant interference with the glucometer test strips and can produce a falsely high reading. [5] In developing countries where there is no new born screening for galactosemia a sick infant with high galactose levels is at risk for developing iatrogenic hypoglycemia. Based on our experience, we suggest precautionary measures to prevent a medical misadventure. All stakeholders including doctors, nurses, diabetic educators and caregivers of galactosemia patients should be made aware of the pitfalls of MutQ-GDH-PQQ chemistry based glucometers. Secondly all glucometers with potential galactose interference should carry prominent warning information. Finally, these glucometers should not be used in a neonatal care setting.

 
   References Top

1.Frias JP, Lim CG, Ellison JM, Montandon CM. Review of adverse events associated with false glucose readings measured by GDH-PQQ-based glucose test strips in the presence of interfering sugars. Diabetes Care 2010;33:728-9.  Back to cited text no. 1
    
2.Available from: http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/PatientAlerts/ucm177189.htm. [Last accessed on 2012 Nov 23].  Back to cited text no. 2
    
3.Rama Devi AR, Naushad SM. Newborn screening in India. Indian J Pediatr 2004;71:157-60.  Back to cited text no. 3
    
4.Clarke SF, Foster JR. A history of blood glucose meters and their role in self-monitoring of diabetes mellitus. Br J Biomed Sci 2012;69:83-9.  Back to cited text no. 4
    
5.Ng WY, Tiong CC, Jacob E. Maltose interference-free test strips for blood glucose testing at point-of-care: A laboratory performance evaluation. Diabetes Technol Ther 2010;12:889-93.  Back to cited text no. 5
    


    Figures

  [Figure 1]
 
 
    Tables

  [Table 1]


This article has been cited by
1 Status of Newborn Screening and Inborn Errors of Metabolism in India
Seema Kapoor,B. K. Thelma
The Indian Journal of Pediatrics. 2018;
[Pubmed] | [DOI]
2 Approach to the interpretation of unexpected laboratory results arising in the care of patients with inborn errors of metabolism (IEM)
Andre Mattman,Murray Potter
Reviews in Endocrine and Metabolic Disorders. 2018;
[Pubmed] | [DOI]
3 Comparative Performance Assessment of Point-of-Care Testing Devices for Measuring Glucose and Ketones at the Patient Bedside
Ferruccio Ceriotti,Ewa Kaczmarek,Elena Guerra,Fabrizio Mastrantonio,Fausto Lucarelli,Francesco Valgimigli,Andrea Mosca
Journal of Diabetes Science and Technology. 2015; 9(2): 268
[Pubmed] | [DOI]



 

Top
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Access Statistics
    Email Alert *
    Add to My List *
* Registration required (free)  

 
  In this article
    Abstract
   Introduction
   Case Report
   Discussion
    References
    Article Figures
    Article Tables

 Article Access Statistics
    Viewed1166    
    Printed28    
    Emailed1    
    PDF Downloaded309    
    Comments [Add]    
    Cited by others 3    

Recommend this journal