Figure 3: System accuracy blood glucose bias plot. Accuracy limit lines are ±15 mg/dl (<100 mg/dl) or ± 15% (≥100 mg/dl). 99.2% (595/600) of glucose results (200 per test lot × 3 lots) (blue dots) fell within the accuracy limit lines. Red dots (5/600) represent results outside the accuracy limit lines. YSI: Yellow Springs Instruments

Figure 3: System accuracy blood glucose bias plot. Accuracy limit lines are ±15 mg/dl (<100 mg/dl) or ± 15% (≥100 mg/dl). 99.2% (595/600) of glucose results (200 per test lot × 3 lots) (blue dots) fell within the accuracy limit lines. Red dots (5/600) represent results outside the accuracy limit lines. YSI: Yellow Springs Instruments